China High Precision Voled Fingertip Oximeter Manufacturer

Display Type: OLED display 
SP02 Measurement Range: 70%-100% 
Accuracy +/-2% on the stage of 70% - 99% 
Pulse Range Measurement Range: 25 – 250 BPM 
Accuracy +/-1BPM or +/-1% (larger)

Quantity:

Display Type: OLED display

SP02 Measurement Range: 70%-100%

Accuracy +/-2% on the stage of 70% - 99%

Pulse Range Measurement Range: 25 – 250 BPM

Accuracy +/-1BPM or +/-1% (larger)

Power Requirement: Two AAA 1.5V alkaline batteries

-           Voltage range: 2.6 - 3.6V

 

Low Consumption: Less than 30mA

Measurement Accuracy: SpO2 Accuracy +/-2% on the stage of 70% - 99%

-           No definition of less than 70%

-           Pulse rate +/-1BPM or +/-1% (larger)

Measurement Performance in Low Perfusion Condition: The pulse wave is available without failure when the simulation pulse wave amplitude is at 6%

 

Anti-interference ability of ambient light: Natural light in indoor lighting and the existing oxygen measurements under dark conditions with the measured values compared to less than 1%.

 

The product will automatically be powered off when no signal is in the product for longer than 8 seconds.

 

Dimension Length: 58 mm, Width: 32mm

-           Weight: 54g (including two AAA batteries)

 

Environment Requirements: Operation Temperature: 40° F - 104°F

                                                Storage Temperature: 50° F - 104°F

 

Problems

Possible reason

Solution

SpO2% or pulse rate can not be shown normally

1. Finger is not plugged correctly

2. Patient’s SpO2value is too low to be measured

1.Retry by plugging the finger

2. There is excessive illumination

3. Measure more times, If you can make sure about no problem existing in the product. Please go to a hospital timely for exact diagnosis

SpO2% or pulse rate is shown unstably

1. Finger might not be plugged deep enough

2. Excessive patient movement

1. Retry by plugging the finger

2. Be calmness.

The Monitor can not be powered on

1.No battery or low power of battery

2.Battery might be installed incorrectly

3. The Monitor might be damaged

1. Please replace battery

2. Please reinstall the battery

3. Please contact with local customer service centre

Indication is suddenly off

1.The Oximeter is automatically powered off  when no signal is detected longer than 8 seconds

2. The batteries power is too low to work

 1. Normal

 2. Replace the battery

“Error3”  or  “Error4” Displayed on screen

1. Err 3 means the red emission LED is damaged.

2. Err 4 means the infra-red emission LED is damaged.

1. Check the red emission LED

2. Check the infra-red emission LED

“Error 6”

Err 6 means the crystal is failure

Change the crystal

“Error7” displayed on screen

Err 7 means all the emission LED or reception dioxide is damaged.

Check the emission LED and reception dioxide.

 

Guidance and Manufacture’s declaration  - electromagnetic emission

Guidance and Manufacture’s declaration – electromagnetic environment specified below. The customer of the user of the Pulse Oximeter should assure that it is used in such and environment.

Emission test

Compliance

Electromagnetic Environment – guidance

RF emissions CISPR 11

Group 1

The pulse Oximeter uses RF energy only for very low and is not likely to cause interference in nearby electronic equipment.

RF emissions CISPR 11

Class B

The pulse Oximeter is suitable for use in all establishments, incluing domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

Harmonic emissions

IEC 61000-3-2

Not Applicable

Voltage fluctuations/ flicker emissions

IEC 61000-3-2

Not Applicable

Previous: 
Next: 

content is empty!

SUNNYWORLD MEDICAL INSTRUMENTS Co., Ltd.
Phone:
+86 577 65917211 / 
66801577 / 66816625
Add:
No.3 , Lane 3, Gao Yang Road, Ruian 325200, Zhejiang, China
ABOUT SUNNYWORLD
We have become one of the leading suppliers of blood pressure monitors, stethoscopes, penlights, diagnostic hammers etc. in China.
LATEST NEWS
SUBSCRIBE​​​​​​​
Sign up for our newsletter to receive the latest news.

​Copyright  2022 RUIAN SUNNYWORLD INTERNATINAL TRADE CO., LTD. Sitemap Support By Leadong.